- New
(edit with the Customer Reassurance module)
(edit with the Customer Reassurance module)
(edit with the Customer Reassurance module)
This part of IEC 62321 specifies two different techniques for the determination of tris(2-chloroethyl) phosphate (TCEP) in plastics, the GC-MS or LC-MS method, both of which are applicable to quantitative analysis.
These two techniques are applicable to use with polyurethane, polyvinylchloride, and polyethylene materials containing TCEP between 200 mg/kg to 2 000 mg/kg.
These test methods do not apply to plastic materials having a processing temperature higher than 230 °C.
GC-MS using a pyrolyser/thermal desorption accessory (Py/TD-GC-MS) technique is described in Annex A and can be used for the screening of TCEP in plastics.
NOTE TCEP starts thermal decomposition at approximately 230 °C. Polymer types that have a processing temperature into shapes of plastics (e.g. pellets, moulded parts or sheets) not exceeding the decomposition temperature can contain TCEP.
See more